  1. To what precision is a standard 50 mL buret read?
    To ±0.01\pm 0.01 mL (two decimal places), estimating one digit between the 0.1 mL markings.
  2. Why is a buret rinsed with a small portion of the titrant before it is filled?
    So that leftover water does not dilute the titrant and change its known concentration.
  3. A calorimeter has a constant of 15 J/°C and its temperature rises by 4.0 °C. How much heat does the calorimeter hardware absorb?
    6060 J, since q=(15 J/°C)(4.0 °C)=60q = (15 \text{ J/°C})(4.0 \text{ °C}) = 60 J.
  4. Heat leaks out of a coffee-cup calorimeter during an exothermic reaction. Is the measured magnitude of ΔH\Delta H too large or too small?
    Too small. Lost heat makes the temperature rise smaller than it should be, so less heat is recorded.
  5. How should the liquid level in a buret be read?
    At the bottom of the meniscus, with the eye level with the meniscus to avoid parallax error.
  6. What criterion is used to choose an indicator for an acid-base titration?
    Its color-change pH range should include the pH at the equivalence point.
  7. At what wavelength should absorbance be measured when building a Beer-Lambert calibration curve?
    At λmax\lambda_{max}, the wavelength of maximum absorbance, which gives the greatest sensitivity and least error.
  8. What does the calorimeter constant (heat capacity of the calorimeter) represent?
    The heat absorbed by the calorimeter itself per degree of temperature rise, in J/°C.
  9. Fingerprints on the optical faces of a cuvette make the calculated concentration of the sample too high or too low?
    Too high. The smudges scatter and absorb light, so the measured absorbance is larger than the analyte alone would give.
  10. Why is a blank measured before samples in spectrophotometry?
    It sets zero absorbance for the solvent and cuvette, so the measured absorbance comes only from the analyte.
  11. An air bubble in the buret tip escapes during a titration. Is the calculated analyte concentration too high or too low?
    Too high. The recorded titrant volume includes the bubble, so it overstates the moles of titrant delivered.
  12. Why does extra distilled water in the titration flask not change the result of a titration?
    It does not change the moles of analyte present, and the endpoint depends on moles, not on volume.