- Identify the correct action if the first drop of capillary blood is used for glucose testing.
- Wipe away the first drop, then collect the next drop per meter IFU. The first drop often contains tissue fluid that can dilute the sample, affecting glucose measurement accuracy.
- Identify the correct action if a point-of-care instrument displays an error code during testing.
- Stop; follow IFU troubleshooting; repeat test only after resolving the error. Error codes indicate instrument issues, and resolution via troubleshooting restores reliability for accurate testing.
- What is the correct action if a QC vial is expired but was used to run controls?
- Invalidate the QC; repeat QC with in-date controls; do not report patient results. Expired controls may not reliably assess test performance due to potential degradation, invalidating the quality check.
- What is the main purpose of quality control (QC) in CLIA-waived testing?
- To verify the test system is functioning properly before reporting results. Quality control confirms the reliability and accuracy of the testing system, preventing erroneous patient results.
- What is the correct action when external controls are out of range or incorrect?
- Do not report patient results; troubleshoot and repeat QC before testing. Incorrect controls signal potential system malfunctions, requiring resolution to maintain result integrity.
- Which two categories of QC are commonly used for waived tests (when applicable)?
- Internal control and external (liquid) control. Internal controls are built into the test device, while external controls use known samples to validate overall test performance.
- Which bloodborne pathogen practice is required when handling blood specimens for waived tests?
- Standard Precautions (treat all blood/body fluids as potentially infectious). Standard Precautions minimize infection risk by assuming all specimens could transmit pathogens.
- Which action is appropriate if a waived test result is inconsistent with patient symptoms?
- Repeat testing and/or refer for confirmatory testing per policy and provider order. Discrepancies may indicate false results, requiring verification to guide appropriate clinical decisions.
- What is the correct disposal method for used lancets and capillary devices?
- Immediately into an approved sharps container. Immediate sharps disposal complies with safety regulations to prevent needlestick injuries and biohazard exposure.
- What is the correct reporting approach for waived test results with qualitative outcomes?
- Report exactly as instructed (for example: positive, negative, invalid). Precise terminology aligns with test validation and prevents miscommunication in clinical reporting.
- What is the correct response if the specimen type does not match the test IFU?
- Reject the specimen and recollect using the required specimen type. Mismatched specimen types can lead to inaccurate results, as tests are validated only for specific sample matrices.
- What does a QC log document for CLIA-waived testing?
- Date/time, test system, lot/exp, QC results, operator, and corrective actions. Comprehensive logging ensures traceability, compliance, and quality assurance in testing processes.
- Which event typically requires running external controls on a waived test (if available)?
- New lot/shipment, new operator, after maintenance, or when results are questionable. These triggers ensure ongoing validation of test accuracy amid changes in materials, personnel, or performance concerns.
- Which step is required when using a glucose meter that requires coding or calibration?
- Verify correct code/calibration for the strip lot before patient testing. Proper calibration aligns the meter with the strip lot's characteristics for accurate result interpretation.
- Which identification elements must be confirmed before performing any waived test?
- Correct patient identity and correct ordered test. Verification of these elements prevents misidentification errors and ensures the test aligns with clinical needs.
- What is the correct site preparation for a fingerstick glucose test before puncture?
- Clean with antiseptic and allow to air-dry completely. Antiseptic cleaning reduces contamination, and air-drying prevents sample dilution or interference.
- Identify the correct action if the internal control line on a rapid test is absent.
- Invalid result; do not report; repeat with a new device and new specimen if needed. Absence of the control line signifies a procedural or device failure, rendering the result unreliable and requiring retesting.
- Which statement best defines an analytic error caused by timing in waived testing?
- Reading results too early or too late relative to the IFU time window. Deviating from the specified timing can alter reaction development, leading to misinterpretation of results.
- Which document is the primary authority for performing a CLIA-waived test correctly?
- The manufacturer's Instructions for Use (package insert/IFU). This document provides authoritative, detailed guidelines from the manufacturer to ensure accurate, safe, and compliant test performance.
- Identify the correct action if a rapid test is read after the IFU maximum read time.
- Do not report; treat as invalid and repeat per IFU with a new device. Results observed outside the validated time frame may reflect unstable reactions, compromising accuracy.
- What does an internal control on a rapid test device primarily indicate?
- The test ran correctly (sample flow/reagent integrity) for that device. Internal controls confirm proper procedural execution, including sample migration and reagent functionality within the specific test device.
- Which action is required before testing if a CLIA-waived kit is past its expiration date?
- Do not use it; discard per policy and obtain an in-date kit. Expired kits contain potentially degraded reagents that could compromise test accuracy and reliability.
- What storage principle must be followed for CLIA-waived reagents and test kits?
- Store exactly as specified in the IFU (temperature, light, humidity, position). Adhering to manufacturer-specified conditions preserves reagent stability and ensures test validity.
- Which practice best prevents cross-contamination during point-of-care waived testing?
- Change gloves as needed and use a new pipette/swab/device for each patient. Using fresh protective and testing equipment per patient minimizes the transfer of contaminants between samples.