Spectrophotometry and Beer–Lambert Law - Biochemistry
- What is the Beer-Lambert law?
- : absorbance equals molar absorptivity times path length times concentration.
- Why are absorbance measurements usually made at ?
- Sensitivity is highest there, and small wavelength errors change the reading least.
- Why is a blank measured before sample absorbance readings?
- It zeroes the instrument for absorbance by the solvent, buffer and cuvette, so readings reflect only the analyte.
- What ratio indicates relatively pure double-stranded DNA?
- About 1.8 (about 2.0 for pure RNA); a lower ratio suggests protein contamination.
- What should be done when a sample's absorbance falls above the linear range of the assay?
- Dilute it into the linear range, measure, and multiply by the dilution factor; high absorbances deviate from Beer-Lambert.
- NADH has Mcm. What is [NADH] if in a 1 cm cuvette?
- M (100 μM). .
- A sample transmits 10% of incident light. What is its absorbance?
- 1.0. .
- Which amino acid residues account for most protein absorbance at 280 nm?
- Tryptophan and tyrosine (with a minor contribution from phenylalanine and disulfides).
- A solution has in a 1 cm cuvette, and Mcm. What is its concentration?
- M (100 μM). .
- If the cuvette path length is doubled and nothing else changes, what happens to absorbance?
- It doubles, because absorbance is directly proportional to path length.
- How is absorbance related to transmittance (as a fraction)?
- .
- Why can many dehydrogenase reactions be followed at 340 nm?
- NADH and NADPH absorb strongly at 340 nm while NAD+ and NADP+ do not, so absorbance tracks cofactor reduction or oxidation.