ARRT Radiography Exam • Procedures

Evaluate Extremity Radiographs

Mastering the systematic evaluation of upper and lower extremity radiographs for diagnostic quality and anatomical accuracy.

A medical scan with bone and radiation symbols.
Inside this lesson 10 sections 5 questions 8 figures

SECTION 1

Historical Context & Motivation

The ability to evaluate extremity radiographs is foundational to the practice of diagnostic radiography, yet the standards and criteria by which we judge image quality have evolved substantially since Wilhelm Conrad Röntgen first captured an X-ray image of his wife's hand in 1895. That seminal image—revealing the bones of the hand and a wedding ring—demonstrated the diagnostic potential of X-rays for skeletal structures, but it also highlighted the immediate need for consistent methods to produce and evaluate such images. As radiographic technology advanced from glass plates to digital detectors, the criteria for evaluating extremity radiographs grew in sophistication, encompassing not only anatomical visualization but also exposure accuracy, positioning precision, and artifact identification.

1895 • Röntgen's Discovery

Wilhelm Röntgen produces the first radiograph of a human hand, immediately demonstrating the value of X-rays for visualizing skeletal extremities and sparking worldwide interest in medical imaging.

1920s • Standardized Positioning Emerges

Pioneering radiographers such as Merrill and Ballinger begin codifying standard projections for the extremities (AP, lateral, oblique), establishing reproducible evaluation criteria that remain in use today.

1970s • Automatic Exposure Control

The introduction of automatic exposure control (AEC) systems reduces exposure variability, but radiographers must still evaluate resulting images for adequate density, contrast, and diagnostic quality.

2000s • Digital Radiography Revolution

Computed radiography (CR) and digital radiography (DR) replace film-screen systems, introducing new evaluation parameters such as exposure index (EI), deviation index (DI), and post-processing algorithms that fundamentally change how technologists assess image quality.

2020s • AI-Assisted Evaluation

Artificial intelligence tools begin assisting radiographers in detecting positioning errors and suggesting repeat criteria, though the technologist's systematic evaluation skills remain the clinical standard.

Throughout this evolution, a central question has persisted: How does a radiographer determine whether an extremity radiograph meets diagnostic standards, and when should an image be repeated? The answer requires a systematic approach that integrates knowledge of anatomy, positioning criteria, exposure parameters, and image quality indicators—skills that the ARRT expects every registered technologist to demonstrate.

SECTION 2

Core Principles of Extremity Radiograph Evaluation

Evaluating an extremity radiograph is not a single judgment but rather a systematic assessment across multiple domains. Each domain addresses a different dimension of image quality and diagnostic utility. The ARRT examination expects candidates to understand and apply these principles consistently across all extremity projections, from the smallest phalanx to the proximal femur. The following core principles form the backbone of any structured evaluation workflow, and mastering them allows the radiographer to make confident decisions about image acceptability in the clinical environment.

1

Anatomical Visualization

Every required anatomical structure must be fully included within the collimation field. For extremity radiographs, this means verifying that the area of interest and at least one adjacent joint are demonstrated without clipping or foreshortening.
2

Positioning Accuracy

Correct positioning produces predictable anatomical relationships. Evaluation criteria include the degree of superimposition of specific structures (e.g., distal radius and ulna overlap on a true lateral wrist), joint space openness, and part rotation indicators.
3

Exposure Quality

Proper exposure ensures adequate visualization of both cortical bone and soft tissue. Technologists assess brightness (exposure index) and contrast resolution, verifying that trabeculae are visible without washout or underexposure.
4

Artifact Identification

Artifacts—such as patient jewelry, clothing, motion blur, grid lines, or processing errors—can obscure anatomy or mimic pathology. The technologist must distinguish clinically significant artifacts from acceptable ones.
5

Marker & Annotation Accuracy

Correct anatomical side markers (R/L), patient identification, and date markers must be present and properly placed. Missing or incorrect markers may necessitate a repeat examination according to department protocol.
✦

KEY TAKEAWAY

Think of evaluating an extremity radiograph like a pilot running through a pre-flight checklist. A pilot doesn't just glance at the cockpit and declare the aircraft ready—every instrument, control surface, and fuel gauge must be verified in sequence. Similarly, a radiographer evaluates each image against a structured checklist of anatomy, positioning, exposure, artifacts, and markers. Skipping a single item can mean the difference between a diagnostic image and one that requires a repeat, adding unnecessary radiation dose to the patient.

SECTION 3

Visual Guide to Extremity Evaluation Criteria

A well-structured evaluation workflow proceeds in a logical order, moving from the broadest assessment to the finest details. The following diagram illustrates the systematic evaluation flowchart that radiographers should follow when reviewing any extremity radiograph. This approach ensures that no critical evaluation criterion is overlooked, regardless of the specific projection or anatomical region being assessed.

EXTREMITY RADIOGRAPH EVALUATION FLOWCHART1. IMAGE RECEIVED2. VERIFY PATIENT ID & MARKERS3. ASSESS ANATOMICAL COVERAGE4. EVALUATE POSITIONING5. CHECK EXPOSURE QUALITY6. IDENTIFY ARTIFACTSACCEPT IMAGESend to PACSREPEAT IMAGEDocument reasonPassFailR/L marker, date,patient nameInclude area ofinterest + 1 jointRotation, tilt,superimpositionEI, DI, contrast,trabecular detail
The evaluation flowchart proceeds from image receipt (Step 1) through six sequential checkpoints. Each step addresses a different quality domain, with annotations on the right indicating the specific criteria assessed. If any criterion fails, the image should be repeated with the reason documented.

As the diagram illustrates, the evaluation process begins with administrative verification—confirming that the correct patient identification and anatomical markers are present—before proceeding to assess anatomical coverage, positioning accuracy, exposure quality, and artifact presence. This sequential approach mirrors clinical workflow and ensures that the most fundamental requirements (correct patient, correct side) are confirmed before investing time in evaluating technical quality. The decision node at the bottom reflects the binary outcome: an image that passes all criteria is sent to the Picture Archiving and Communication System (PACS), while a failed image must be repeated with the specific reason documented for quality assurance tracking.

SECTION 4

Exposure Parameters & Quality Indicators

While extremity radiograph evaluation is primarily qualitative, digital radiography has introduced quantitative metrics that guide the technologist's assessment of exposure adequacy. Understanding these metrics and their mathematical foundations is essential for the ARRT examination and for clinical practice in any digital imaging department. The two most important values in digital radiography are the Exposure Index (EI) and the Deviation Index (DI), both standardized by the International Electrotechnical Commission (IEC).

EXPOSURE INDEX (IEC STANDARD)

EI=c×g(Dˉ)EI = c × g(D̄)

Where EI is the exposure index, c is a calibration constant specific to the detector system, and g(D̄) is a function of the median detector dose value in the region of interest. A higher EI indicates greater detector exposure.

DEVIATION INDEX

DI=10×log10(EI/EIT)DI = 10 × log₁₀(EI / EI_T)

Where DI is the deviation index, EI is the actual exposure index of the image, and EI_T is the target exposure index for that specific examination. A DI of 0 indicates perfect exposure. Acceptable range is typically −1.0 to +1.0; values outside −3.0 to +3.0 generally require a repeat.

The deviation index is the more clinically actionable metric because it directly communicates how far the actual exposure deviates from the ideal target. A positive DI indicates overexposure (the detector received more radiation than necessary), while a negative DI indicates underexposure. Although digital systems can compensate for moderate exposure errors through post-processing, the ALARA principle demands that technologists strive for a DI as close to zero as possible. Overexposure delivers unnecessary radiation dose to the patient, while underexposure increases quantum noise and may obscure subtle pathology such as hairline fractures.

SPATIAL RESOLUTION (LP/mm)

R=1/(2×pixelpitch)R = 1 / (2 × pixel pitch)

Where R is the limiting spatial resolution in line pairs per millimeter (lp/mm), and pixel pitch is the center-to-center distance between adjacent detector elements in mm. Extremity radiography demands high spatial resolution (typically ≥ 5 lp/mm) to visualize fine trabecular detail and cortical margins.

⚠

CLINICAL NOTE

Some vendors use proprietary exposure indicators (e.g., Fuji's S-number, Carestream's EI) that do not follow the IEC standard. In these systems, the relationship between indicator value and exposure may be inversely proportional (as with Fuji's S-number, where a higher number means less exposure). Always verify which system your clinical site uses and know its acceptable ranges.

SECTION 5

Positioning Evaluation by Anatomical Region

Each extremity projection has specific positioning evaluation criteria that serve as benchmarks for determining whether the projection was performed correctly. These criteria are derived from the expected anatomical relationships when the body part is aligned with the image receptor (IR) and central ray (CR) according to standard protocols. The following diagram and table summarize critical criteria for the most commonly tested extremity projections on the ARRT examination.

POSITIONING CRITERIA: KEY EXTREMITY PROJECTIONSHAND — PA✓ All phalanges included✓ Fingers slightly separated✓ No rotation (equalconcavity of MC shafts)✓ Distal radius/ulnaincluded✓ CR to 3rd MCP jointWRIST — LATERAL✓ Ulnar styloid in profile✓ Radius/ulnasuperimposed distally✓ Pronator fat padvisible✓ Metacarpalssuperimposed✓ CR to midcarpal areaKNEE — AP✓ Open femorotibialjoint space✓ Fibular head slightlysuperimposed on tibia✓ Patella centered overfemoral condyles✓ Femoral condylessymmetric in shapeANKLE — AP (MORTISE)✓ Open ankle mortise(15−20° internal rotation)✓ Tibial plafond visible✓ Equal joint space onmedial & lateral sides✓ Distal fibula free oftibial superimpositionELBOW — LATERAL✓ Elbow flexed 90°✓ Olecranon in profile✓ Three concentric arcs(trochlear sulcus,capitellum, trochlea)✓ Anterior/posteriorfat pads evaluatedFOOT — AP (DORSOPLANTAR)✓ All phalanges &metatarsals included✓ CR angled 10° towardheel (cephalad)✓ Open MTP joint spaces✓ Tarsals at leastpartially included
Six commonly tested extremity projections with their key positioning evaluation criteria. Each card highlights the anatomical landmarks, superimposition patterns, and central ray placements that the ARRT expects candidates to recognize and assess on both AP/PA and lateral projections.
ProjectionRotation IndicatorWhat Rotation Means
Hand PAUnequal concavity of metacarpal shafts; phalanges asymmetricHand was not flat on the IR; part was rotated medially or laterally
Knee APFibular head overlap with tibia: too much = external rotation; too little = internal rotationLeg was rotated from true AP position; affects femorotibial joint space visualization
Elbow LateralSeparation of the three concentric arcs (trochlear sulcus, capitellum, trochlear notch)Humerus was not parallel to IR or elbow not flexed to 90°
Ankle MortiseUnequal medial and lateral clear spaces of the mortise jointInsufficient or excessive internal rotation from the standard 15−20°

Common rotation indicators and their clinical significance for frequently tested extremity projections

SECTION 6

Worked Example: Evaluating an AP Knee Radiograph

Consider a clinical scenario in which you have just acquired an AP knee radiograph on a 45-year-old patient with an order indicating 'right knee pain, rule out fracture.' The image is now displayed on your digital workstation. Walk through the systematic evaluation process as follows.

Systematic Evaluation of an AP Knee Radiograph

  1. 1

    Step 1 — Verify Patient ID & Markers

    Confirm that the patient's name, date of birth, medical record number, and date of examination are correctly displayed on the image. Verify that a right (R) anatomical marker is visible on the image and does not superimpose any anatomy of interest. The marker should be placed outside the collimation field or in a non-diagnostic area.
    ✓ Pass — R marker present, ID confirmed
  2. 2

    Step 2 — Assess Anatomical Coverage

    For an AP knee, the image must include approximately 3–4 inches (7–10 cm) of the distal femur and proximal tibia/fibula. The femorotibial joint space must be fully demonstrated. Check that the collimation is tight enough to minimize scatter but wide enough to include all required anatomy. Soft tissue margins should be visible on both medial and lateral aspects of the knee.
    ✓ Pass — Adequate coverage with both distal femur and proximal tibia visible
  3. 3

    Step 3 — Evaluate Positioning

    Examine the relationship of the fibular head to the proximal tibia. On a properly positioned AP knee, approximately one-half of the fibular head should be superimposed by the lateral tibial condyle. Check that the patella is centered between the femoral condyles, and that the femoral condyles appear relatively symmetric in size and shape. If the medial condyle appears significantly larger, the leg was externally rotated; if the lateral condyle appears larger, it was internally rotated.
    ⚠ Minor finding — Fibular head shows slightly excessive overlap, suggesting 5–10° of external rotation. Evaluate if joint space is still diagnostically open.
  4. 4

    Step 4 — Check Exposure Quality

    Review the exposure index (EI) and deviation index (DI) displayed by the system. For this image, EI = 320 and EIT = 250. Calculate the DI: DI = 10 × log10(320/250) = 10 × log10(1.28) = 10 × 0.107 ≈ +1.07. This is slightly above the ideal range of −1.0 to +1.0, indicating mild overexposure. The trabecular pattern of the distal femur should still be visible; verify cortical bone margins and soft tissues are well demonstrated.
    ⚠ DI = +1.07 — Slightly overexposed but within acceptable limits (< +3.0). Reduce mAs for subsequent projections.
  5. 5

    Step 5 — Identify Artifacts & Final Decision

    Scan the entire image for artifacts. Confirm there are no metallic objects (jewelry, snaps, zippers), no motion blur at the cortical margins, and no processing artifacts such as dead pixel lines or detector element dropout. In this case, the image is free of artifacts. Given that the positioning showed only minor external rotation and the femorotibial joint space remains diagnostically open, the image meets acceptability criteria.
    ✓ ACCEPT — Image is diagnostic. Note mild overexposure for technique adjustment. No repeat needed.
ℹ

REPEAT CRITERIA REMINDER

An image should be repeated when: (1) the area of clinical interest is not fully demonstrated, (2) rotation or positioning errors prevent visualization of the joint space or critical anatomy, (3) the DI exceeds ±3.0, (4) motion blur obscures cortical margins, or (5) artifacts superimpose the area of interest. When in doubt, consult the supervising radiologist before repeating—unnecessary repeats add patient dose without clinical benefit.

SECTION 7

Common Errors & Corrective Actions

Understanding common errors in extremity radiography and their corrective actions is essential for both clinical practice and ARRT examination success. The following table categorizes the most frequently encountered errors, their radiographic appearances, and the appropriate corrective measures. Recognizing these patterns allows the technologist to efficiently diagnose the cause of a suboptimal image and take corrective action without unnecessary repeated exposures.

Error TypeRadiographic AppearanceCorrective Action
Part rotationAsymmetric joint spaces, altered superimposition patterns, foreshortened or elongated structuresReposition the part to eliminate rotation; use immobilization devices (sponges, sandbags) to maintain position
Motion blurBlurred cortical margins, double contour lines on trabecular pattern, loss of spatial resolutionUse shortest possible exposure time; immobilize part; reassure patient; consider using a higher mA with lower time
Clipped anatomyRequired structures (e.g., adjacent joint, distal phalanx) are cut off at the collimation borderReposition the part centrally on the IR; adjust collimation or use a larger IR; ensure CR is directed to the correct centering point
Overexposure (high DI)DI significantly positive (> +1.0); image appears acceptable on screen but delivers unnecessary dose; may lose contrast in thin-tissue areasReduce mAs by approximately 30% for a 1-stop correction; ensure collimation is tight (excessive field size drives up EI)
Underexposure (low DI)DI significantly negative (< −1.0); increased quantum mottle (noise); grainy appearance especially in dense anatomyIncrease mAs; verify correct body part selection in the exposure technique chart; ensure the generator is functioning properly
Foreign body artifactRadiopaque objects (jewelry, splints, clothing snaps) superimposing anatomy of interestRemove all removable objects before exposure; if non-removable (surgical hardware), document and angle around if possible

Common extremity radiography errors with their radiographic appearances and corrective actions

✦

KEY TAKEAWAY

Think of common radiographic errors as a troubleshooting decision tree, similar to how an automotive mechanic uses a diagnostic flow when a check-engine light appears. The symptom (blurred image, clipped anatomy, noisy appearance) points you to a specific root cause (motion, misalignment, underexposure), and each root cause has a defined corrective action. Over time, experienced technologists develop pattern recognition that allows near-instant identification of the problem—much like an experienced mechanic who can identify an engine misfire by sound alone.

SECTION 8

Advanced Evaluation: Pathology, Pediatrics, & Trauma

The standard evaluation criteria for extremity radiographs serve as a baseline, but clinical reality frequently demands modifications for special populations and circumstances. Trauma patients may not be able to achieve standard positioning due to pain, splinting, or fracture displacement; pediatric patients present with unfused epiphyseal plates that must be distinguished from fracture lines; and patients with known pathology (e.g., osteoporosis, rheumatoid arthritis, Paget's disease) may require adjusted exposure techniques and modified evaluation criteria. Understanding these advanced considerations distinguishes the competent technologist from the exceptional one.

Standard EvaluationAdvanced / Modified Evaluation
Exact positioning to standard criteria (e.g., 90° flexion for lateral elbow)Trauma modifications: cross-table lateral, modified positioning to accommodate splints/casts; document adaptation
Adult skeletal anatomy: fused epiphyses, predictable cortical patternsPediatric evaluation: identify and differentiate growth plates from fractures; use comparison views of contralateral extremity when indicated
Standard exposure techniques based on body part measurementsPathology-adjusted technique: decrease for osteoporotic bone, increase for casted extremities (typically 2× mAs for fiberglass, 3−4× for plaster)
Single-view evaluation with standard criteriaRule of twos: minimum two views at 90° apart; some departments require two joints to be demonstrated for long bone injuries
DI within ±1.0 considered idealWider DI tolerance may be accepted for trauma if positioning constraints prevent optimal technique; clinical context matters

Comparison of standard and advanced evaluation criteria for special clinical scenarios

As you advance in your career—and potentially pursue certifications in CT, MRI, or post-primary specializations—the analytical framework you develop for evaluating extremity radiographs will extend directly into cross-sectional imaging. The same systematic approach applies: verify patient information, confirm anatomical coverage, assess image quality, identify artifacts, and determine diagnostic adequacy. Mastering this framework for radiographs establishes the cognitive habits that will serve you in every advanced modality.

SECTION 9

Practice Problems

Question 1: CONCEPTUAL

A radiographer evaluates a PA hand radiograph and notices that the metacarpal shafts show unequal concavity—the second and third metacarpals appear more concave on one side than on the other. What positioning error does this finding indicate, and how does it affect diagnostic quality?
Show answer
Unequal concavity of the metacarpal shafts on a PA hand radiograph is the hallmark indicator of part rotation. In a properly positioned PA hand, both sides of each metacarpal shaft should demonstrate equal, symmetric concavity because the hand is flat against the image receptor with no tilt. When one side appears more concave, the hand was rotated either medially or laterally, which causes obliquity of the metacarpals and phalanges. This rotation can foreshorten or elongate the joints, potentially closing interphalangeal joint spaces and obscuring subtle fractures or joint pathology. If the rotation is significant enough to affect joint space visualization, the image should be repeated with careful attention to placing the hand flat on the IR.

Question 2: BASIC CALCULATION

A digital radiograph of the wrist produces an EI of 400. The target EI (EIT) for this examination is 200. Calculate the deviation index (DI) and determine whether the image should be repeated based on the DI alone.
Show answer
Using the formula DI = 10 × log₁₀(EI / EI_T), we calculate DI = 10 × log₁₀(400 / 200) = 10 × log₁₀(2.0) = 10 × 0.301 = +3.01. A DI of +3.01 indicates that the detector received approximately twice the intended radiation dose. This value falls at the upper boundary of the generally accepted range (±3.0). Most department protocols would flag this as requiring corrective action—either a repeat or at minimum a documented technique adjustment for subsequent images. The overexposure delivers unnecessary radiation dose to the patient in violation of the ALARA principle, even though the image may appear visually acceptable on the monitor due to digital post-processing compensation.

Question 3: INTERMEDIATE

A lateral elbow radiograph demonstrates clear separation of the three concentric arcs (trochlear sulcus, capitellum, and trochlear notch), and the olecranon process does not appear in true profile. Additionally, the anterior fat pad is elevated and a posterior fat pad is visible. Evaluate this image for positioning accuracy and clinical significance.
Show answer
This image demonstrates a positioning error and a clinically significant finding. The separation of the three concentric arcs and the olecranon not being in true profile indicate that the elbow was rotated—likely the distal humerus was not parallel to the image receptor, or the elbow was not flexed to exactly 90°. On a true lateral elbow, the trochlear sulcus, capitellum, and trochlear notch should form three tightly concentric arcs. However, the elevated anterior fat pad and the visible posterior fat pad constitute the radiographic 'fat pad sign,' which is highly suggestive of an intra-articular effusion, most commonly caused by an occult fracture (especially a radial head fracture). Despite the positioning error, if the joint space and fat pads are adequately demonstrated, the radiologist may still be able to make a clinical diagnosis. The technologist should note the positioning limitation and consult with the radiologist before repeating, as the fat pad finding may be the most diagnostically important feature.

Question 4: APPLIED

A 7-year-old child presents to the emergency department with a swollen right wrist after a fall. The PA wrist radiograph is obtained and shows radiolucent lines at the distal radius and distal ulna that could represent either normal physis (growth plate) or fracture lines. The DI is −0.5. What evaluation criteria should the radiographer apply, and what additional actions should be considered?
Show answer
In pediatric patients, the presence of open physes (growth plates) creates radiolucent lines that can mimic fracture lines, making evaluation more complex than in adults. The radiographer should first confirm that the DI of −0.5 is within acceptable range (it is—well within ±1.0), ensuring adequate image quality. The evaluation should then focus on whether the radiolucent lines are smooth, well-defined, and bilaterally symmetric (characteristic of normal physes) versus irregular, angulated, or associated with cortical buckling (suggestive of a Salter-Harris fracture). The technologist should ensure that both the proximal and distal joints are included, as pediatric fractures can occur at growth plate sites distant from the point of maximal tenderness. A comparison view of the contralateral (left) wrist should be considered per department protocol, as it provides a normal baseline for that individual child's skeletal development. The radiographer should communicate any concerns about the growth plate appearances to the radiologist at the time of image submission.

Question 5: CRITICAL THINKING

A technologist obtains an AP ankle radiograph that demonstrates the following: the medial clear space of the ankle mortise measures 5 mm while the lateral clear space measures only 2 mm; the DI is +0.3; the anatomical marker is correctly placed; and the distal tibia and fibula are both fully included. The technologist accepts the image as diagnostic. Evaluate this decision and provide a comprehensive analysis of whether this image meets evaluation criteria.
Show answer
The technologist's decision to accept this image warrants careful scrutiny. While the DI of +0.3 is excellent, the marker is correct, and anatomy is fully included, the significant asymmetry in the medial (5 mm) versus lateral (2 mm) clear spaces of the ankle mortise represents either a positioning error or a pathological finding—both of which demand attention. In a properly positioned AP mortise view (15−20° internal rotation), the medial and lateral clear spaces should be approximately equal. An asymmetry of this magnitude could indicate: (1) insufficient internal rotation, causing the lateral malleolus to superimpose the tibial plafond and artificially narrowing the lateral clear space; or (2) a true pathological finding such as a deltoid ligament injury with medial clear space widening, which would be clinically critical in the setting of a Weber B or C fibular fracture. The technologist should not simply accept the image but should first verify whether 15−20° of internal rotation was applied. If positioning was correct and the asymmetry persists, the finding is potentially pathological and must be immediately communicated to the radiologist. If the positioning was incorrect, a repeat with proper rotation is indicated. This scenario demonstrates why positioning evaluation cannot be divorced from clinical context—what appears to be a positioning error may actually be a critical diagnostic finding.

SUMMARY

Summary: Evaluating Extremity Radiographs

Summary

Evaluating extremity radiographs requires a systematic, sequential approach that addresses five critical domains: patient identification and markers, anatomical coverage (including the area of interest and at least one adjacent joint), positioning accuracy (verified by superimposition patterns, rotation indicators, and joint space openness), exposure quality (assessed through the deviation index, targeting DI values within ±1.0), and artifact identification. Each projection has specific evaluation criteria—such as metacarpal concavity for hand rotation, fibular head overlap for knee rotation, and concentric arcs for lateral elbow positioning—that must be memorized and applied consistently.

Beyond standard evaluation, the competent radiographer adapts their assessment for trauma, pediatric, and pathological conditions, recognizing that modified positioning may be necessary and that certain findings (such as fat pad signs or growth plate variations) carry clinical significance that extends beyond image quality. The deviation index (DI = 10 × log₁₀(EI / EI_T)) provides a quantitative framework for exposure assessment, while the ALARA principle demands that every unnecessary repeat be avoided. Mastering this evaluation framework is not merely an academic exercise for the ARRT examination—it is the daily clinical standard that ensures every patient receives diagnostic-quality images with minimal radiation exposure.

Varsity Tutors • ARRT Radiography Exam • Evaluate Extremity Radiographs