Recognize Equipment Malfunctions - ARRT Radiography Exam
- Which immediate action is required if the exposure switch sticks in the ON position during an exposure?
- Release the switch and use the emergency-off/power disconnect immediately. This action prevents prolonged radiation exposure, ensuring patient and operator safety by immediately halting the malfunctioning exposure process.
- What is the appropriate action if the Bucky tray does not lock securely in place?
- Do not expose; remove from service and report mechanical failure. An unsecured tray risks movement artifacts or misalignment during exposure, compromising image quality and patient safety.
- Which indicator most strongly suggests a failing x-ray tube filament?
- Intermittent exposures or repeated filament error messages. These signs reflect filament degradation, which impairs electron emission and leads to unreliable x-ray production.
- Identify the appropriate action if a portable x-ray unit will not hold a charge or loses power quickly.
- Remove from service and report battery/charging system failure. Power issues compromise reliable operation during mobile procedures, requiring service to restore functionality and prevent interruptions.
- Which action is required if a ceiling-suspended tube crane drifts or will not hold position?
- Remove from service and report mechanical counterbalance/brake failure. Drifting indicates faulty brakes or counterbalances, posing injury risks from uncontrolled movement during positioning.
- Identify the likely cause when CR images show a consistent wide white band along one edge.
- CR reader transport/scan error; clean per policy and report if unresolved. The band artifact results from incomplete scanning or debris interference, warranting cleaning to restore uniform image processing.
- What is the correct action when a CR reader repeatedly jams or fails to read plates?
- Stop processing, remove from service, and contact service support. Repeated failures disrupt workflow and may damage plates, requiring service to diagnose and repair mechanical or optical components.
- Which response is appropriate if the exposure indicator or technique display is blank or flickering?
- Do not expose; power down if needed and report control console failure. Display malfunctions prevent accurate technique selection, risking improper exposures and necessitating repair for reliable operation.
- What is the correct response if the AEC terminates exposure too late, causing overexposure across exams?
- Discontinue AEC use and report for physics/service calibration. Systematic overexposure indicates a global calibration fault in automatic exposure control, mandating manual techniques for dose control.
- Which malfunction is suggested by a burning odor or smoke from the x-ray tube housing?
- Electrical overheating or arcing; shut down and remove from service. These symptoms indicate potential electrical failure or insulation breakdown, necessitating immediate removal to avoid fire hazards or equipment damage.
- What is the correct action if an error message indicates a failed system self-test at startup?
- Do not use clinically; follow vendor restart steps and report to service. Failed self-tests signal underlying system issues that could affect performance, mandating vendor protocols to ensure safe operation.
- Which reporting step is appropriate after identifying a malfunction that could affect patient safety or dose?
- Document, tag out of service, notify supervisor, and contact service/physics. This protocol ensures traceability, prevents use of faulty equipment, and facilitates prompt expert intervention for safety and compliance.
- Identify the likely malfunction when grid motion is absent on a table Bucky and grid lines appear.
- Reciprocating grid failure; remove from service and report for repair. Lack of motion in a reciprocating grid causes visible lines due to stationary attenuation, indicating a mechanical failure needing repair.
- Which corrective action is appropriate when the collimator lamp is out and a replacement is available?
- Replace the collimator bulb per policy and verify light field operation. Replacement restores visibility for accurate collimation, with verification ensuring proper alignment and function before resuming use.
- Which malfunction is most likely when a DR image shows repeating dead pixels or lines in the same location?
- Detector element failure; perform QC/bad pixel map and report if persistent. Persistent defects in fixed positions indicate hardware failure in the detector array, requiring quality control to assess and mitigate impact.
- What does repeated arcing with loud snapping sounds during exposure preparation suggest?
- High-voltage arcing; shut down and remove the system from service. Arcing noises during prep phase indicate high-voltage breakdown, posing risks of tube failure or electrical hazards.
- Identify the most likely cause when the collimator light fails to illuminate but exposure is possible.
- Burned-out collimator lamp or failed collimator power supply. The light operates independently from the x-ray generation system, so failure points to issues specific to the bulb or its power source.
- What is the proper response if a DR detector shows visible cracks or fluid intrusion is suspected?
- Remove detector from use, isolate it, and report immediately. Physical damage can lead to electrical shorts or image artifacts, necessitating isolation to prevent further degradation or safety risks.
- What is the correct response if a tube rotor continues to spin after the exposure is completed?
- Remove the unit from service and report for service evaluation. Continued spinning indicates a control circuit failure that could lead to tube damage, requiring professional evaluation to prevent further issues.
- Which problem is suggested when the AEC consistently produces images that are too dark for one chamber?
- AEC chamber or calibration error; stop using that chamber and report. Consistent underexposure from one chamber points to localized detector or calibration issues, requiring isolation to avoid diagnostic errors.
- What should be done if the interlock system fails and exposure is possible with the door open?
- Stop use immediately, secure the room, and report a critical safety failure. Interlock failure bypasses essential radiation safety protocols, requiring immediate securing to prevent unauthorized exposure risks.
- Which response is required if the tube head or collimator housing is unusually hot to the touch?
- Stop use, allow cooling, and report possible overheating malfunction. Excessive heat suggests cooling system failure, which can cause thermal damage to components if operation continues.
- Which issue is indicated if the PBL (automatic collimation) selects an incorrect field size repeatedly?
- PBL sensor/calibration malfunction; disable PBL and report for service. Repeated errors suggest sensor or software faults in the positive beam limitation system, necessitating manual mode to maintain accuracy.
- Identify the appropriate response if a lead apron shows cracks on inspection or fails fluoroscopic check.
- Remove from clinical use and tag for replacement per policy. Cracks compromise radiation shielding effectiveness, requiring replacement to maintain personnel protection during procedures.
- What does a light field that does not match the x-ray field indicate during a collimation check?
- Collimator misalignment; remove from service and report for calibration. Mismatch signifies calibration errors that could result in improper radiation fields, increasing patient dose and requiring recalibration.