Evaluate Digital Image Characteristics - ARRT Radiography Exam
- Which change increases spatial resolution: smaller pixels or larger pixels?
- Smaller pixels increase spatial resolution. Reducing pixel size allows for finer sampling of spatial frequencies, enhancing the ability to resolve small objects.
- Which sampling condition prevents aliasing: sampling at or ?
- Sampling at (Nyquist criterion). The Nyquist criterion ensures accurate reconstruction by sampling at least twice the highest frequency, preventing foldover artifacts.
- What is aliasing in digital imaging?
- False pattern from undersampling high spatial frequencies. Aliasing occurs when spatial frequencies exceed the Nyquist limit, causing high-frequency signals to appear as lower-frequency artifacts.
- Which change increases spatial resolution: larger matrix or smaller matrix (same FOV)?
- Larger matrix increases spatial resolution. Increasing matrix size for a fixed FOV reduces pixel size, improving the system's ability to capture higher spatial frequencies.
- What is spatial resolution in digital radiography?
- Ability to display small objects as separate (recorded detail). It quantifies the system's capability to distinguish fine anatomical details by resolving closely spaced structures without blurring.
- What is signal-to-noise ratio (SNR) in digital imaging?
- Ratio of useful signal magnitude to noise magnitude. SNR measures image quality by comparing the strength of the desired signal to background noise fluctuations.
- What is dynamic range in digital radiography?
- Range of exposures producing useful image receptor response. Dynamic range defines the detector's linear response to varying X-ray intensities, enabling capture of a wide exposure latitude.
- Which characteristic is primarily improved by greater bit depth: contrast resolution or spatial resolution?
- Contrast resolution. Greater bit depth increases the number of gray levels, enhancing the detection of subtle attenuation differences without affecting pixel size.
- What is modulation transfer function (MTF) used to describe in digital imaging?
- Spatial resolution performance versus spatial frequency. MTF quantifies how effectively the imaging system preserves contrast as spatial frequency increases, indicating resolution efficiency.
- How many gray levels are available with a 12-bit system?
- gray levels. A 12-bit system provides 4096 discrete values, enabling finer gradations in image intensity for improved contrast.
- What does wide dynamic range allow regarding exposure variation?
- Acceptable images over a broad range of exposures. A wide dynamic range provides exposure latitude, compensating for variations while maintaining diagnostic image quality.
- Which change generally increases SNR: higher mAs or lower mAs (all else constant)?
- Higher mAs increases SNR (more photons, less quantum mottle). Higher mAs increases photon fluence, reducing quantum noise and improving the signal relative to noise fluctuations.
- What is contrast resolution in digital imaging?
- Ability to display small differences in tissue attenuation. Contrast resolution enables visualization of subtle density differences, limited by the system's ability to detect small signal variations.
- What is bit depth in digital imaging?
- Number of bits used to represent each pixel value. Bit depth determines the precision of pixel value quantization, allowing representation of subtle intensity differences.
- State the formula for the number of gray levels from bit depth.
- . The number of gray levels exponentially increases with bit depth, as each bit doubles the possible intensity values.
- State the formula for SNR in terms of signal and noise.
- . SNR is calculated as the ratio of average signal intensity to the standard deviation of noise, indicating image clarity.
- State the formula for pixel size using field of view and matrix size.
- . Pixel size is derived by dividing the field of view by the matrix dimension, determining the spatial sampling interval.
- For quantum-limited imaging, how does SNR change with the number of detected photons ?
- . In quantum-limited systems, SNR improves with the square root of photon count due to Poisson statistics reducing relative noise.
- What is exposure creep in digital radiography?
- Gradual increase in technique due to wide dynamic range masking overexposure. Exposure creep results from technicians incrementally raising exposure factors, as digital systems tolerate overexposure without visible saturation.
- How many gray levels are available with a 10-bit system?
- gray levels. A 10-bit system yields 1024 intensity levels, sufficient for many imaging applications but less than higher bit depths.
- Which change increases spatial resolution: smaller FOV or larger FOV (same matrix)?
- Smaller FOV increases spatial resolution. Decreasing FOV for a fixed matrix reduces pixel size, enabling better resolution of fine details within the imaged area.
- State the formula for Nyquist frequency in terms of pixel size.
- . The Nyquist frequency represents the maximum spatial frequency that can be accurately sampled without aliasing, half the reciprocal of pixel size.
- What is the limiting spatial resolution primarily determined by in DR/CR detectors?
- Detector element size (pixel size or phosphor laser spot size). Smaller detector elements capture finer details, directly limiting the maximum resolvable spatial frequency in digital systems.
- What is quantum mottle in digital radiography?
- Noise from insufficient photon numbers (quantum noise). Quantum mottle arises from statistical variations in low photon counts, manifesting as graininess in underexposed images.
- If detected photons increase from to , how does SNR change (quantum-limited)?
- SNR doubles because . Quadrupling photons increases SNR by a factor of two, as noise scales with the square root of photon count.